System Messages

System Messages

System messages remain on the display until the specified action is taken or no longer relevant. Each message is accompanied by a three second beep to alert you. Table 12-1 lists system messages.

Table 12-1 System Messages

Message

Description

Corrective Action

 

 

 

$WWDFK￿3DGV￿&DEOH

The pads cable is not properly attached to

Check the cable connection.

 

the device.

 

 

 

 

&RQILJXUDWLRQ￿/RVW

The configuration is reset to the default

￿Reconfigure the HeartStart XLT.

 

settings.

If the message persists, call for ser-

 

 

vice.

 

 

 

'DWD￿&DUG￿'LVDEOHG

The PC card is not in use because it is

If possible, turn the HeartStart XLT off

 

full, incompatible, absent, or inserted

for more than 2 minutes, insert an empty,

 

after the HeartStart XLT was turned on.

compatible Data Card, and turn the device

 

 

on.

 

 

 

(&*￿)DXOW

The ECG data acquisition system failed

Remove the device from active use and

 

and data is unavailable from the 3- or 5-

call for service.

 

lead monitoring electrodes.

 

 

 

 

/RZ￿%DWWHU\

The battery has sufficient capacity

Replace the battery with a fully charged

 

remaining to provide only about ten min-

battery.

 

utes of monitoring time and six shocks

 

 

before the HeartStart XLT shuts off.

 

 

 

 

/HDGV￿2II

The monitoring electrodes are not

Check the monitoring electrodes are

 

applied.

￿properly applied.

 

The monitoring electrodes are not

Check the monitoring cable is prop-

 

making proper contact with the

erly connected.

 

￿patient.

 

 

The monitoring cable is not con-

 

 

nected.

 

 

 

 