Operational Checks

Power Supply

Battery - make sure:

a charged battery is in the HeartStart XLT

￿another battery is charged or being charged the batteries have no visible signs of damage

AC/Power Module

1.Make sure a battery is in the HeartStart XLT.

2.Plug the power module into a power outlet and connect it to the Heart- Start XLT.

3.Verify that the power and charging indicators on the power module are lit.

4.Remove the battery from the HeartStart XLT and verify that the charg- ing indicator on the power module is no longer lit. Replace the battery.

Printer - make sure the printer:

￿has sufficient paper prints properly

11-6

Maintaining the HeartStart XLT

Page 120
Image 120
Philips M3500B manual Power Supply