Operational Checks

 

 

 

 

 

The report also lists additional checks that you should do. Perform each of

 

 

these checks and record the results. The guidelines for completing the checks

 

 

are as follows:

11

 

 

Defibrillator Inspection - make sure the HeartStart XLT is clean, clear of

 

 

objects on top and has no visible signs of damage.

 

Cables/Connectors - make sure there are no cracks, broken wires, or other visible signs of damage. Make sure the connectors engage securely.

Supplies - make sure the carrying case has:

two sets of multifunction defib electrode pads in sealed packages, within the expiration date

an adequate supply of monitoring electrodes, within the expiration date

alcohol wipes hand towel scissors

a razor

an extra roll of printer paper a spare charged battery

￿a Data Card

SpO2 sensors (if monitoring SpO2)

M3500B HeartStart XLT Defibrillator/Monitor

11-5

Page 119
Image 119
Philips manual M3500B HeartStart XLT Defibrillator/Monitor 11-5