PrintStrip .
Print
Strip
PrintStrip .
Print Strip

Printing Events

Printing Events

The HeartStart XLT can be configured to print automatically when certain events occur. The table below lists these events and the length of the strip printed, depending on whether the printer is configured to print real-time or with a 6-second delay.

Table 9-4 Printing Events

 

Event

Real-Time

Delayed

 

Strip Length

Strip Length

 

 

 

 

 

Defibrillator charges

continuous

6 seconds just prior to charging, plus

 

 

 

continuous printing throughout the

 

 

 

charge duration.

 

 

 

Shock Delivered

12 seconds

6 seconds just prior to shock, plus 12

 

 

 

seconds after shock.

 

 

 

Shock Failed

6 seconds

6 seconds just prior to the message

 

 

 

1R￿6KRFN￿'HOLYHUHG, plus 6 seconds after

 

 

 

the message.

 

 

 

Defibrillator disarmed

6 seconds

6 seconds just prior to shock, plus 6

 

 

 

seconds after shock.

 

 

 

Alarm Violation

6 seconds

6 seconds just prior to alarm viola-

 

 

 

tion, plus 6 seconds after alarm vio-

 

 

 

lation.

 

 

 

 

Mark

pressed

6 seconds

6 seconds just prior to marking, plus

Event

 

6 seconds after button pressed.

 

 

 

 

 

 

 

Printing is configured independently for each of these events. You can stop

the printing before it has printed the entire strip by pressing .

To print additional events that you observe in the course of caring for your

patient, press An ECG strip will print continuously until you press

a second time to stop printing. If the printer is configured to have a 6- second delay, the printout contains an additional 6 seconds of the ECG that occurred just prior to pressing

9-8

Storing, Retrieving & Printing

Page 98
Image 98