Pacing

Pacing

To perform pacing:

1. Press Pacer . The green LED next to Pacer lights up and the dia- logue box appears.

Figure 8-1 Pacing Display

HR ALARM

78

150 J

LEAD

SELECT

Lead II

Pacer Stop

Demand Mode 70ppm 30mA

SpO2

Shocks:0

00:00:8

ENERGY CHARGE

SELECT

The message 6WRS indicates that the pacing function is on but paced pulses are not being delivered. The pacer turns on in the mode last used.

2.Verify that the dot markers appears near the middle of the QRS complexes of the ECG.

If the dot markers do not appear, or are in the wrong location, adjust the ECG size or select another lead. (See “Monitoring the ECG” in Chapter 4.)

The ECG will only appear if the 3- or 5-lead monitoring cable and elec- trodes are used.

3. Press Mode to change to fixed mode, if R-wave detection is unreliable. The pacing dialogue box displays the current mode. To switch back to demand mode, press Mode again.

8-4

Pacing

Page 86
Image 86
Philips M3500B manual Pacing, Pacer Stop Demand Mode 70ppm 30mA