Pacing
8-4 Pacing
Pacing
To perform pacing:
1. Press . The green LED next to lights up and the dia-
logue box appears.
Figure 8-1 Pacing Display
The message 6WRS indicates that the pacing function is on but paced pulses
are not being delivered. The pacer turns on in the mode last used.
2. Verify that the dot markers appears near t he middle of the QRS compl exes
of the ECG.
If the dot markers do not appear, or are in the wrong location, adjust the
ECG size or select another lead. (See “Monitoring the ECG” in Chapter
4.)
The ECG will only appear if the 3- or 5-lead monitoring cable and elec-
trodes are used.
3. Press to change to fixed mode, if R-wave detection is unrel iable .
The pacing dialogue box displays the current mode. To switch back to
demand mode, press again.
Pacer
Pacer
ENERGY
SELECT
HR ALARM LEAD
SELECT
CHARGE
Shocks:0
00:00:8
78 Lead II
SpO2
150 J
Pacer Stop
Demand Mode
70ppm 30mA
Mode
Mode