System Messages

 

 

 

 

 

Table 12-1 System Messages (Continued)

 

 

 

 

 

 

 

 

Message

Description

Corrective Action

 

 

 

 

 

 

 

1R￿3DGV

The multifunction defib electrode pads

Check the pads cable connection.

 

 

 

are not properly connected to the Heart-

 

 

 

 

Start XLT.

 

 

 

 

 

 

 

 

3DGV￿&DEOH￿2II

The pads cable is not connected to the

Check pads connector is locked in place.

 

 

 

defibrillator.

 

 

 

 

 

 

 

 

 

 

 

 

3DGV￿2II

The pads are not making proper contact

Make sure the pads are properly applied

 

12

 

with the patient.

to the patient.

 

 

 

 

 

 

3DFHU￿)DLOXUH

The pacing system is not functioning.

Remove the device from active use and

 

 

 

 

call for service.

 

 

 

 

 

 

 

3DFHU￿2XWSXW￿/RZ

High patient impedance is resulting in the

Check the pads are applied properly.

 

 

 

pacer delivering less current to the patient

 

 

 

 

than specified in the output current set-

 

 

 

 

ting.

 

 

 

 

 

 

 

 

6HUYLFH￿8QLW

Appears during a Shift/System Check.

Replace the Data Card.

 

 

 

May indicate that the Data Card is full.

￿Perform a Shift/System Check

 

 

 

OR

 

 

 

 

There is a system failure.

If 6HUYLFH￿8QLW continues to appear, do

 

 

 

 

not use the device and call for service.

 

 

 

 

 

 

 

6\VWHP￿)DLOXUH￿&\FOH￿3RZHU

An internal error occurred.

If this occurs during actual use:

 

 

 

 

Substitute another defib, if possible.

 

 

 

 

Remove this unit from clinical use and

 

 

 

 

￿call for service.

 

 

 

 

If no other defib is available, turn

 

 

 

 

power off, then on. If unit turns on

 

 

 

 

normally, use for this one incident.

 

 

 

 

After this incident, remove from clini-

 

 

 

 

￿cal use and call for service.

 

 

 

 

If unit does not turn on normally, it

 

 

 

 

cannot be used.

 

 

 

 

If this occurs during routine testing:

 

 

 

 

Remove this unit from clinical use and

 

 

 

 

call for service.

 

 

 

 

 

 

 

M3500B HeartStart XLT Defibrillator/Monitor

12-3

Page 139
Image 139
Philips M3500B manual 3DGV&DEOH2II