System Messages

Table 12-1 System Messages (Continued)

Message

Description

Corrective Action

 

 

 

6S2￿￿&DEOH￿2II

The SpO2 cable is not connected to the

Attach the SpO2 cable to the HeartStart

 

device.

XLT.

 

 

 

632￿￿)DLOXUH

A failure has occurred in the SPO2

Remove the device from active use and

 

circuitry.

call for service.

 

 

 

6S2￿￿/LJKW￿,QWHUI

The level of ambient light is so high that

Cover the sensor with an opaque

 

the sensor cannot obtain an SpO2 reading

material.

 

or the sensor or cable is damaged.

Check the sensor for damage; try

 

 

another sensor.

 

 

 

6S2￿￿1RQ￿3XOVDWLOH

The patient’s pulse is absent or too weak

Check that the sensor is applied prop-

 

to be detected.

erly.

 

 

Make sure the sensor site has a pulse.

 

 

Relocate the sensor to a site with

 

 

improved circulation.

 

 

Try another sensor.

 

 

 

6S2 ￿/RZ￿6LJQDO

SpO2 signal is too low to give an accurate

Check the sensor is applied properly.

 

reading.

Try another sensor type.

 

 

 

6S2￿￿1RLV\￿6LJQDO

Excessive patient movement, electrical

Minimize patient movement or apply

 

interference, or optical interference is

the sensor to a site with less move-

 

present.

ment.

 

 

Secure the sensor cable loosely to the

 

 

￿patient.

 

 

Reduce sources of electrical or optical

 

 

interference.

 

 

 

6S2￿￿6HQVRU￿)DLO

The SpO2 transducer is broken.

Try another sensor.

 

 

 

12-4