Troubleshooting

Troubleshooting

The table below lists the pacing-related system and momentary messages that you may encounter during pacing.

Table 8-1 Pacing System Messages

 

 

 

 

 

 

 

 

 

Message

 

Possible Cause

Corrective Action

 

 

 

 

 

 

 

/HDGV￿2II

The selected monitoring

Check that the monitoring

 

 

 

lead is not making proper

electrodes are properly

 

 

 

￿contact with the patient.

￿applied.

 

 

 

Pacing was attempted in

Check that the monitoring

 

 

 

demand mode without

cable and electrodes are

 

 

 

monitoring electrodes

properly connected.

 

 

 

 

 

attached.

 

 

 

 

 

 

 

8

￿3DFHU￿)DLOXUH

The pacing system is not func-

Remove the device from active

 

 

tioning.

 

use and call for service.

 

 

 

 

 

 

 

3DFHU￿2XWSXW￿

High patient impedance is

Check that the pads are applied

 

 

/RZ

resulting in the delivery of less

properly.

 

 

 

current to the patient than

 

 

 

 

specified in the output current

 

 

 

 

setting.

 

 

 

 

 

 

 

 

 

 

6WRS￿3DFHU￿

Mode

is pressed while

Stop pacing before changing

 

 

 

the pacing mode.

 

 

 

paced pulses are being deliv-

 

 

 

 

 

 

 

ered.

 

 

 

 

 

 

 

 

 

 

.H\￿,QDFWLYH

Pacer

, or one of the other

Make sure Manual Mode is

 

 

 

active before pressing

 

 

 

pacing function keys, is

 

 

 

Pacer or one of the other

 

 

 

pressed when Manual Mode is

 

 

 

 

 

 

 

not active.

pacing function keys.

 

 

 

 

 

 

 

 

M3500B HeartStart XLT Defibrillator/Monitor

8-7

Page 89
Image 89
Philips M3500B manual Pacing System Messages, Message Possible Cause Corrective Action