Troubleshooting
M3500B HeartStart XLT Defibrillator/Monitor 8-7
8
Troubleshooting
The table below lists the pacing-related system and momentary mes sages that you may encounter during pacing.
Table 8-1 Pacing System Messages
Message Possible Cause Corrective Action
/HDGV2II The selected monitoring
lead is not making proper
contact with the patient.
Pacing was attempted in
demand mode without
monitoring electrodes
attached.
Check that the monitoring
electrodes are properly
applied.
Check that the monitoring
cable and electrodes are
properly connected.
3DFHU)DLOXUH The pacing system is not func-
tioning. Remove the device from active
use and call for service.
3DFHU2XWSXW/RZ
High patient impedance is
resulting in the delivery of less
current to the patient than
specified in the output current
setting.
Check that the pads are applied
properly.
6WRS3DFHU is pressed while
paced pulses are being deliv-
ered.
Stop pacing before changing
the pacing mode.
.H\,QD FWLYH , or one of the other
pacing function keys, is
pressed when Manual Mode is
not active.
Make sure Manual Mode is
active before pressing
or one of the other
pacing function keys.