Troubleshooting Tips

Table 12-3 Troubleshooting Tips (Continued)

Situation

Cause

Possible Solution

 

 

 

There is a dashed (----)

ECG data is not being acquired.

Check the patient cable is connected.

line on the display

 

Check the pads or electrodes are prop-

instead of an ECG.

 

erly applied.

 

 

Check that the desired lead is selected.

 

 

 

The HeartStart XLT

￿The battery is low.

￿Change the battery.

does not appear to be

There is a system failure.

Take the device out of use and call for

functioning properly.

 

service.

 

 

 

The displayed time is

The time was not correctly set in the con-

Set the time in the General Settings menu

incorrect.

figuration.

of the Configuration Mode.

 

 

 

The printed date is

The date was not correctly set in the con-

Set the time in the General Settings menu

incorrect.

figuration.

of the Configuration Mode.

 

 

 

12-8

Troubleshooting

Page 144