Specifications
ECG Monitoring
Inputs: Single channel ECG may be viewed on display and printed. PADS ECG is obtained through two multifunction defibrillation electrode pads. Lead I, II, or III is obtained through the
Lead Fault: LEADS OFF message and dashed line appear on the monitor if a lead electrode or wire becomes disconnected.
Pad Fault: PADS OFF message and dashed line appear on the monitor if a pad becomes disconnected.
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an accuracy of + 10%.
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm lim-
its: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Defibrillation Patient Cable Length: 7 ft (2.13 m).
ECG Cable Length: 10 ft (3.05 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI stan- dards for cardiac monitors (EC 13).
ECG Size: 2.5, 5, 10, 20, 40 mm/mV.
Frequency Response:
AC Line Filter: | 60 Hz/50Hz. |
Pads ECG for Display: Monitor
Pads ECG for Printer: Monitor
Leads ECG for Display: Monitor
Leads ECG for Printer: Diag
Monitor
Patient Isolation (defibrillation proof):
ECG: Type CF
SpO2: Type CF
Defib: Type BF
Specifications & Safety |