Monitoring
M3500B HeartStart XLT Defibrillator/Monitor 5-7
5
Monitoring
To monitor SpO2:
1. If the HeartStart XLT is not on, press On.
2. Apply the appropriate sensor to the patient.
3. Make sure the sensor cable is connected to the HeartStart XLT.
4. Press to turn on SpO2 monitoring.
A dashed line (---) is displayed under , while the oxygen saturation is
measured and an SpO2 value is calculated. In a few seconds the SpO2 value is
displayed in place of the dashed line. As the patient’s oxygen saturation
changes, the SpO2 value is updated continuously.
Figure 5-3 SpO2 Monitoring Display
To the right of the SpO2 value, a pleth bar and SpO2 alarm indicator are dis-
played. The pleth bar should be observed for fluctuation. It is an indication of
pulsation detected by the sensor. The pleth bar should not be used as the sole
indicator of pulsation because it can be influenced by movement and artifact.
The symbol indicates no alarm is set.
Below the SpO2 value is the pulse rate derived from the pulse oximetry.
6S2
6S2
PAUSE
HR ALARM LEAD
SELECT
ANALYZE
Shocks: 3
00:00:49
Monitoring Rhythm
78 Pads
SpO2
98
PR 62