To monitor SpO2:

1.If the HeartStart XLT is not on, press On.

2.Apply the appropriate sensor to the patient.

3.Make sure the sensor cable is connected to the HeartStart XLT.

4. Press 6S2￿to turn on SpO2 monitoring.

A dashed line (---) is displayed under , while the oxygen saturation is measured and an SpO2 value is calculated. In a few seconds the SpO2 value is displayed in place of the dashed line. As the patient’s oxygen saturation changes, the SpO2 value is updated continuously.

Figure 5-3 SpO2 Monitoring Display

LEAD

HR ALARMSpO2

SELECT

78

 

Pads

98

PR 62

Monitoring Rhythm

PAUSE

Shocks: 3 00:00:49

ANALYZE

To the right of the SpO2 value, a pleth bar and SpO2 alarm indicator are dis-

played. The pleth bar should be observed for fluctuation. It is an indication of

 

pulsation detected by the sensor. The pleth bar should not be used as the sole

5

indicator of pulsation because it can be influenced by movement and artifact.

The

symbol indicates no alarm is set.

 

Below the SpO2 value is the pulse rate derived from the pulse oximetry.