Operational Checks

Shift/System Check

To perform the Shift/System Check:

1.Turn the HeartStart XLT off.

2.Connect a 50 ohm test load to the pads patient cable (instead of pads).

3.If a Data Card is routinely used, insert a Data Card into the HeartStart XLT.

4.If a power module is used, unplug the power module.

5.Insert a charged battery.

6.While pressing PrintStrip , press On to start the test.

7.Follow the prompts on the display to proceed with the test. If the message 6HUYLFH￿8QLW appears, check the information in Table 12-2 and Table 12-3 .

If the message continues to appear, do not use the device, and call for service.

The test takes less than a minute to complete. When it is done, a report is printed, as shown in Figure 11-1. Test results are listed as Pass or Fail, with the exception of the Data Card Test. For the Data Card Test, the results indi- cate either the amount of storage time remaining on the data card or that no data card is present.

Figure 11-1 Shift/System Check Report

Shift/System Check

8 Jan 1999 13:52:17

SN:US00000001

Current Tests:

 

Qty/Check List:

 

___ Defibrillator Inspection

General System Test

Pass

___ Cables/Connectors

ECG Test

Pass

___ Defibrillation Pads

Backup Power Test

Pass

___ Monitoring Electrodes

SpO2 Test:

Pass

___ Charged Batteries

Data Card Test

2:00

___ AC/DC Power Module

Defib Test

Pass

___ Printer Paper

Pacer Test

Pass

___ Data Card

 

 

 

 

___ Ancillary Supplies

 

 

___ SpO2 Sensor

11-4

Maintaining the HeartStart XLT