Troubleshooting

 

 

 

 

Table 5-2 System Messages (Continued)

 

 

 

 

 

 

Problem or Message

Possible Cause

Corrective Action

 

 

 

 

 

6S2￿&DEOH￿2II

The SpO2 cable is not

Attach the cable to the

 

 

connected to the device.

HeartStart XLT.

 

 

 

 

 

6S2￿￿6HQVRU￿)DLO￿

The transducer is bro-

Apply a new trans-

 

 

ken.

ducer.

 

 

 

 

 

6S2￿)DLOXUH￿

￿There is a hardware fail-

￿Remove device from

 

 

ure in the SpO2 sub-

active use and call for

 

 

system.

service.

 

 

 

 