Monitoring Rhythm

If Rhythm Monitoring detects a shockable rhythm, you are prompted as fol- lows:

Figure 3-13 Shockable Rhythm

Check Patient

If No Pulse

Press ANALYZE Shocks: 3

00:00:55

PAUSE ANALYZE

This prompt is repeated periodically, as configured, until

$1$/<=(

or

 

is

3$86(

 

 

 

 

 

 

3

pressed. If you press

 

, the defibrillation process starts again.

$1$/<=(

If you press , rhythm monitoring is suspended for the duration of the

pause. is used when administering CPR, as noted earlier. It may also be useful when performing medical procedures or encountering artifact during patient transport. Active SpO2 and heart rate alarms are suspended during the pause duration, as well.

Press 5(680( to restore rhythm monitoring. Active SpO2 and heart rate alarms are also restored.

M3500B HeartStart XLT Defibrillator/Monitor

3-15