Getting Acquainted

Figure 2-3 AED Mode Display Layout (ECG and SpO2 disabled)￿

System Message

Momentary Message

ECG

User Message

ECG Gain

150J Current Charge

Shocks: 0 Shocks Delivered

00:15:02 Incident Timer

Defibrillation

 

PAUSE

ANALYZE

SHOCK

Soft Keys

 

 

 

 

 

 

 

 

 

 

NOTE

 

In its default configuration, the HeartStart XLT powers on in AED Mode

 

 

where voice prompts and user messages are active.

 

 

 

 

 

 

 

The Incident Timer shows the elapsed time since the HeartStart XLT was

 

 

turned on. If the HeartStart XLT is powered on after being off for less than

 

 

two minutes, the Incident Timer resumes where it left off. If power is off for

 

 

more than two minutes, the Incident Timer resets to zero (￿￿￿￿￿￿￿￿).