Applying Monitoring Electrodes

Figure 4-2 Precordial Lead Electrode Placement

 

 

 

 

Electrode

Location

 

 

 

 

 

 

 

 

 

 

 

V1

C1

forth intercostal space, at right sternal

 

 

 

 

 

 

margin

 

 

 

 

 

 

 

 

 

 

 

V2

C2

forth intercostal space, at left sternal

1

2

 

 

 

 

margin

 

 

 

 

 

 

3

5

 

V3

C3

midway between V2/C2 and V4/C4

 

4

6

 

 

 

 

V4

C4

fifth intercostal space, at left midclav-

 

 

 

 

 

 

 

 

 

 

icular line

 

 

 

 

 

 

 

 

 

 

 

V5

C5

same level as V4/C4, on anterior axil-

 

 

 

 

 

 

lary line

 

 

 

 

 

 

 

 

 

 

 

V6

C6

same level as V4/C4, at left mid axil-

 

 

 

 

 

 

lary line

 

 

 

 

 

 

 

Table 4-2 5-Lead ECG Lead Formation

Lead

 

 

 

 

Lead Formation

 

 

 

 

 

 

 

 

I

LA - RA

 

 

 

 

 

 

 

 

 

 

 

 

II

LL - RA

 

 

 

 

 

 

 

 

 

 

 

 

III

LL - LA

 

 

 

 

 

 

 

 

 

 

 

 

aVR

RA -

 

LA + LL

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

aVF

LL -

RA + LA

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

aVL

LA -

RA + LL

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Vx (or Cx)

 

 

RA + LA + LL

 

V/C -

 

 

 

 

 

where x = 1-6

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 