Philips M3500B manual Specifications & Safety

Models: M3500B

1 180
Download 180 pages 16.23 Kb
Page 166
Image 166

Safety Considerations

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

WARNING

Do not allow multifunction defib electrode pads to touch each other or to touch other ECG monitoring electrodes, lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and patient skin burns during defibrillation and may divert current away from the heart.

During defibrillation, air pockets between the skin and multifunction defib electrode pads may cause patient skin burns. To help prevent air pockets, make sure the pads completely adhere to the skin. Do not use dried out pads; do not open pads package until just prior to use.

Never touch the patient or any equipment connected to the patient (including the bed or gurney) during defibrillation.

Do not operate the HeartStart XLT in standing water. When using the HeartStart XLT in wet environments, make sure the Data Card door is securely shut.

Do not immerse, or pour fluids on, any portion of the HeartStart XLT.

Do not use the HeartStart XLT in a flammable or oxygen-rich atmosphere. This can cause an explosion hazard.

Avoid connecting the patient to several devices at once. Leakage current limits may be exceeded. Do not use a second defibrillator on the patient while pacing with the HeartStart XLT.

13-20

Specifications & Safety

Page 166
Image 166
Philips M3500B manual Specifications & Safety