Cleaning Instructions
M3500B HeartStart XLT Defibrillator/Monitor 11-13
11
Cleaning the Printer Printhead
If the printout has light or varying density printing, clean the printhead to
remove any buildup of paper residue.
To clean the printhead:
1. Slide the printer door to the right until the paper roller pops up.
2. Pull up on the plastic removal tab to remove the roll of paper.
3. Clean the printhead surface (above the brush) with a cotton swab dipped
in rubbing alcohol.
4. Replace the roll of paper (see “Loading Printer Paper” on page 11-10).
Cleaning the Power Modules
The following cleaning products may be used to clean the exterior surfaces of
the AC power modules:
Isopropyl alcohol (70% in water)
Mild soap and water
Chlorine bleach (3% in water)
Quaternary ammonium compounds, such as Lysol (10% in water)
When cleaning:
Avoid pouring fluids on the module. Do not allow fluids to penetrate
the exterior surfaces of the module.
Do not use abrasive cleaners or strong solvents such as acetone or
acetone-based compounds.