Specifications

Defibrillator (cont.)

AED Mode

AED Energy Protocol: Fixed Energy (150 Joules).

AED Shock Series: 2, 3, or 4.

Shock Series Timer: off, 30, 60, 90, 120, 150, 180, or 210 seconds.

Text and Voice Prompts: Extensive text audible messages guide user through protocol.

Protocol: Follows pre-configured settings. Can be configured to meet American Heart Association and European Resuscitation Council guide- lines.

AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock, Lead Select, SpO2, HR Alarms, Display Contrast, ECG Gain, Volume, Print Strip, Print Summary, Annotated Mark Event, Manual Mode Entry.

Indicators: LCD Display for ECG waveform and text prompts, Audio

Alerts, Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone, 13 Printer.

Armed Indicators: Charge Done Tone, Available Energy indicated on display, Voice Message.

Patient Analysis: Per protocol, evaluates patient ECG and signal quality to determine if a shock is appropriate and evaluates connection impedance for proper defibrillation pad contact.

Shockable Rhythms: Ventricular fibrillation with amplitude > 100 uV and wide complex ventricular tachycardia with rates greater than 150 bpm.

Sensitivity and Specificity: Meets AAMI guidelines.