STP-H600/H1000 Series Instruction Manual

15.3Abnormalities When Performing The STP Pump Start Operation

(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any Abnormality/Error Warning Lamp Lights.")

Table 15.2 Troubleshooting When Performing The STP Pump Start Operation

No.

Symptom

Probable Cause

Countermeasures

Referred Section

The "ACCELERATION"

(1)

TB3

When this function

8.2

 

lamp does not light.

 

INHIBIT (5) -

is not used, short the

 

 

 

 

(7) for

circuit between

 

 

 

 

Rotation

terminals (5) - (7).

 

 

 

 

INHIBIT

If these terminals

 

 

 

 

Signal Input

are opened, the STP

 

 

 

 

are opened.

pump will not rotate

 

 

 

 

 

after performing

 

 

 

 

 

start operation.

 

 

 

(2)

The

Slide the "MANUAL/

4.1.2

 

 

 

"MANUAL/R

REMOTE"

5.5

 

 

 

EMOTE"

changeover switch

5.6

 

 

 

changeover

on the rear panel of

 

 

 

 

switch is not

the STP control unit

 

 

 

 

in the correct

to the correct

 

 

 

 

position.

position.

 

 

 

(3)

Blown fuses

Replace the blown

4.1.3

 

 

 

F11, 12 (only

fuses (250 V, 0.1 A).

8.4

 

 

 

REMOTE

 

12.3

 

 

 

operation).

 

 

15.4Abnormalities While the STP Pump is Rotating

(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any Abnormality/Error Warning Lamp Lights.")

Table 15.3 Troubleshooting While the STP Pump Is Rotating

No.

Symptom

 

Probable Causes

Countermeasures

Referred

 

Section

 

 

 

 

 

1

The "NORMAL

(1)

Insufficient pressure

Use the maximum

3.3.4

 

OPERATION

 

at the inlet port.

working pressure or less.

16.1

 

"lamp does not

(2)

Insufficient pressure

Use the allowable backing

3.3.4

 

light approx. 15

 

at the outlet port.

pressure or less.

16.1

 

minutes after

(3)

Leakage of the piping

Check piping.

3.3.4

 

performing start

 

system.

 

 

 

operation.

(4)

Failure in the

Check the auxiliary pump

3.3

 

 

 

auxiliary pump.

for its capacity and

16.1

 

 

 

 

START state.

 

 

 

 

 

(use a pump with the

 

 

 

 

 

recommended capacity or

 

 

 

 

 

more).

 

 

 

(5)

Incorrect connection of

Connect the motor

4.4.2

 

 

 

the motor connection

connection cable correctly.

 

 

 

 

cable.

If the cable connector is

 

 

 

 

 

not locked, lock it

 

 

 

 

 

securely.

 

15-3