3.2Precautions Before Installation

STP-H600/H1000 Series Instruction Manual

3.2Precautions Before Installation

3.2.1Operating Environment

Manual backgroundManual background CAUTION Manual background

Chlorine or fluorine system gases can be used in chemical specific pumps (type C). When you use gases including alkaline metals, but excluding Li, gases including Ga, Hg, Sn, or HBr, contact Seiko Instruments.

NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the STP-H600/STP-H1000 pump or other models without anti-corrosion treatment (see Section 1.1, "Usable Gases").

Install the STP pump in a place meeting the following requirements:

Ambient Temperature

0 to 40 °C

Ambient Relative Humidity

30 to 95 % (no dew condensing)

Environment

￿￿ A place free of externally-applied mechanical shock.

 

￿￿ A place free of a heat source

 

(Keep clean of the heat source or attach a thermal

 

shield plate).

 

￿￿ A place free of a strong magnetic field

 

(Range: up to 15mT (150G) in the axial direction,

 

and up to 3mT (30G) in the radial direction with

 

respect to the rotational axis of the STP pump).

 

￿￿ A place free of a strong electric field.

 

￿￿ A place free of exposure to radiation.

 

￿￿ No discharge of high voltage (more than 500 V)

 

(If more than 500 V is discharged, contact Seiko

 

Instruments).

STP Pump Installation

￿￿ Install the STP pump securely so that foreign

Equipment Conditions

particles will easily fall into the STP pump (Ex.: Si

 

wafers or samples are positioned above the STP

 

pump) (To prevent foreign particles from falling into

 

the STP pump, design a shield plate with large

 

conductance).

3-3

Page 25
Image 25