3)Other Cases:

STP-H600/H1000 Series Instruction Manual

2)In case the FREE RUN state is not reset after troubleshooting:

i.Check the tachometer on the front panel.

ii.If the needle is in the red, it means the STP pump has stopped. Power OFF and switch the breaker "OFF."

iii.If the needle is in the black, the STP pump is rotating.

Wait for it to enter the red. Then, power OFF and switch the breaker "OFF. "

iv.Restart the STP pump and check if operates correctly.

Manual backgroundManual background CAUTION Manual background

DO NOT power OFF and switch the breaker OFF until the needle is in the red.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

If the FREE RUN state cannot be reset, it takes approx. 8 to 10 hours to stop the STP pump (the needle is in the red) because it rotates by inertia.

To stop the STP pump quickly, close the vacuum valve at the outlet port flange and introduce gas from the purge port into the STP pump.

3)Other Cases:

After the STP pump stops thoroughly, power OFF, switch the breaker "OFF" and remove probable causes of the abnormality/error. Then, restart the STP pump and check if it operates correctly.

15.2Abnormalities When Powering ON

(When any "Abnormality/Error Warning" lamp lights, see Section 15.5 "When Any Abnormality/Error Warning Lamp Lights.")

Table 15.1 Troubleshooting After Powering ON

No.

Symptom

Probable Cause

Countermeasures

Referred Section

The "POWER ON/OFF"

(1)

Incorrect

Connect the power

4.4.2

 

switch built-in lamp

 

connection of

cable correctly.

 

 

(green LED) does not

 

the power

 

 

 

light, when powering

 

cable.

 

 

 

ON.

 

 

 

 

 

 

(2)

Blown fuses

Replace blown fuses

4.1.3

 

 

 

F3, F4.

(250 V, 5 A).

12.3

 

 

(3)

Abnormal

Set the voltage to the

4.4.2

 

 

 

power voltage

rated voltage ￿10%.

16.2

 

 

(4)

Power

Check whether or

6.1.1

 

 

 

failure.

not a power failure

 

 

 

 

 

has occurred.

 

15-2