2)Connecting the STP Connection Cable

Connect the socket side of the STP connection cable to the STP connector of the STP pump and connect the pin side to "P.CONNECTOR CON5A" of the STP control unit.

Manual backgroundManual background CAUTION Manual background

Always use the STP pump, STP control unit and the STP connection cables of the same model name, serial number and cable length. Failure to do so may result in product damage.

Contact your nearest Seiko Seiki representative if you plan to use units with the same model name but different serial numbers and cable length; They must be adjusted. In some cases, the configuration may need more adjustment.

Use the STP connection cable that has a label affixed

STP-H600/H1000

Series

The use of different cables may result in product damage.

3)Connecting the Motor Connection Cable

Connect the socket side of the motor connection cable to the motor connector of the STP pump and connect the pin side to "INVERTER OUTPUT CON2" of the STP control unit.

Manual backgroundManual background CAUTION Manual background

Use the motor connection cable that has a label affixed

STP-H600/H1000

Series

 

 

The use of different cables may result in product damage.

4)Connecting the Emergency Vent. Valve Cable

Connect the emergency vent. valve to the purge port of the STP pump and the connector to "LEAK VALVE CON8A" of the STP control unit (see Section 3.3.5, "Connecting the Emergency Vent. Valve").

4-14