DC I/O TB3 :

STP-H600/H1000 Series Instruction Manual

DC I/O TB3 :

Remote Input Signal Terminal Block

+12V

1 k Ω

1

3

1 k Ω

N.O : Normal Open

N.C : Normal Close

IN : Input Terminal

OUT : Output Terminal

R.P.THER

Auxiliary pump

 

thermal switch

N.C IN

input terminal

5

INHIBIT

Rotation inhibit signal

 

 

7

N.C IN

input terminal

 

1 k Ω

HEATING

 

2

Baking heater control

 

 

4

N.O IN

signal input terminal

 

9

L..VALVE

Emergency vent. valve

C R 8

 

 

open signal output terminal

10

N.O OUT

 

6

 

Ground terminal

8

 

Unused terminal

Manual backgroundNOTICEManual backgroundManual backgroundManual background

Short the circuit between (1)-(3), when the R.P.THER terminal is not used.

Short the circuit between (5)-(7), when the INHIBIT terminal is not used.

See Table 8.2 for the contact ratings of relay CR8.

Figure 8.4 DC I/O TB3 Remote Input Signal Terminal Block