3 Installation of the STP pump

STP-H600/H1000 Series Instruction Manual

3 Installation of the STP pump

3.1Name and Function of Each Part

(1)Inlet Port Flange (ICF*1 , VG*2 , ASA, ISO, etc.)

￿￿Connected to the vacuum equipment (at the high vacuum side).

3.1Name and Function of Each PartNEVER remove it. CAUTION Manual background

A splinter shield is attached to the inlet port flange to prevent foreign particles from falling into the STP pump.

NEVER remove it.

(2)Outlet Port Flange (KF*2 40)

￿￿Connected to the inlet port side of the auxiliary pump.

(3)STP Connector (41 pins)

￿￿Connected to the STP connection cable.

(4)Motor Connector (5 pins)

￿￿Connected to the motor connection cable.

(5)Purge Port (KF*2 10)

￿￿Introduces a purge gas.

In order to protect the inside of the STP pump when sucking reactive or corrosive gases.

The STP pump is delivered with a blank flange attached to this port.

(6)Cooling Water Port (PT*2 (R/C)1/4 Female Screw)

￿￿Connected to the STP pump cooling water pipe. This port is used when water cooling the STP pump.

(7)Emergency Vent. Valve

￿￿Protects the STP pump.

It functions immediately after any abnormality/error occurs inside the STP pump.

(8)Ground Terminal

￿￿Used for grounding.

Connect the ground cable between this terminal and the ground terminal of the STP control unit .

The ground terminal is marked with label Manual background.

(9)Temperature Sensor Connector (optional accessory)

￿￿A temperature sensor is attached.

For use with the Temperature Management System (TMS) unit only.

*1 : JVIS

*2 : JIS

3-1

Page 23