4.4.2How to Connect the Cables

STP-H600/H1000 Series Instruction Manual

4.4.2How to Connect the Cables

Manual backgroundManual background CAUTION Manual background

When connecting/disconnecting cables, always power OFF the STP pump (switch the breaker "OFF").

Failure to do so may result in electric shock or product damage.

With each connector, align the position of the guide key and insert the pin vertically so as not to bend it.

If the pin is bent, not only may the connector not function normally, but may the pin make contact, resulting in a malfunction.

Lock and securely tighten each connector and screw.

DO NOT apply voltage to each pin and DO NOT cause any short-circuiting between pins.

Connect each cable securely. DO NOT place heavy objects on the cables nor bend them excessively.

Support each cable so as not to apply direct force to the connectors or terminals. If any problem occurs in cables, connectors or terminals, the STP pump may not function normally.

1)Connecting Ground Cable

Connect the ground cable (yellow/green) between the ground terminal of the STP pump and the ground terminal of the STP control unit.

When the resistance between the ground terminals is lower than 0.1 ￿, it is not necessary to connect the ground cable after installing the STP pump and the STP control unit.

Manual backgroundManual background CAUTION Manual background

First, connect the ground cable. Next, connect other cables.

When the resistance between the ground terminals is over 0.1 ￿, always connect the ground cable. Failure to do so may result in electric shock.

4-13

Page 50
Image 50
Seiko Instruments MT-17E-003-D instruction manual 4.4.2How to Connect the Cables