10.3How to Charge the External Battery

STP-H600/H1000 Series Instruction Manual

10.3How to Charge the External Battery

Manual backgroundNOTICEManual backgroundManual backgroundManual background

The battery is charged while the power is "ON," regardless of the operation state of the STP pump.

The STP control unit is equipped with a constant voltage (55 V) circuit and a constant current (0.3 A) circuit for charging the battery.

After a power failure, charge the battery for approx. 12 times the BATTERY BACKUP time. For example, if the BATTERY BACKUP time is 6 minutes, it takes approx. 75 minutes to charge the battery.

If a power failure occurs 2 times within 90 minutes, charge the battery for more than 90 minutes before starting the STP pump.

10.4How to Dispose of the External Battery

Dispose of or collect worn-out batteries in accordance with the guidelines given by each national and/or local government. If you procure the batteries, dispose of them according to the guidelines given by the battery manufacturer.

Manual backgroundManual background CAUTION Manual background

Pay attention to short-circuiting or leakage of the batteries when disposing of them.

To prevent the batteries from short-circuiting, remove the lead wires and connectors from the batteries.

Cut the lead wires one by one so as not to cause a short-circuiting.

Coat the conductive part of the batteries with tape.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

It is recommended to dispose of worn-out batteries after discharging them thoroughly.

Batteries manufactured by Japan Storage Battery Co., Ltd. will be collected by Japan Storage Battery Co., Ltd. For detailed information, contact Japan Storage Battery Co., Ltd. or Seiko Seiki.