5.6.4Powering OFF

STP-H600/H1000 Series Instruction Manual

5.6.4Powering OFF

When all three ("ACCELERATION", "NORMAL OPERATION" and "BRAKE") lamps are off, switch the breaker OFF on the rear panel.

The magnetic bearing stops, and the rotor lands. The POWER ON/OFF switch built-in lamp goes out.

Manual backgroundManual background CAUTION Manual background

NEVER switch OFF the breaker on the rear panel while any of the "ACCELERATION," "NORMAL OPERATION" and "BRAKE" lamps is being lit. If doing so under the above condition, the STP pump may react as if there was a power failure.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

The brake of the STP pump disengages at approx. 1000 rpm and the "BRAKE" lamp goes out. The STP pump rotates by inertia even after the "BRAKE" lamp has gone out.

Powering OFF immediately after the "BRAKE" lamp goes out causes the touch down bearing to touch the rotor resulting in a noise.

Frequent POWER OFF operations may expedite the wear of the touch down bearing. It is recommended to power OFF the STP pump after it stops thoroughly.