STP-H600/H1000 Series Instruction Manual

PROCEDU

Lamp Being Lit

Probable Causes

Countermeasures

Referred

RE No.

 

 

 

Section

No.4

“FAILURE”

(3) Overheating inside the

 

 

 

 

STP control unit:

1) Set the ambient

4.2.1

 

 

1) Ambient temperature.

 

 

 

temperature to 40 ￿C or

 

 

 

2) Insufficient cooling of

lower.

4.2.2

 

 

2) Remove obstacle from the

 

 

the STP control unit.

ventilation port.

12.2

 

 

 

If the air cooling fan is

 

 

 

broken, contact Seiko Seiki.

 

 

 

(4) Peripheral equipment

Remove abnormalities from

8.2

 

 

abnormality (when

peripheral equipment.

 

 

 

peripheral equipment is

 

 

 

 

connected between TB3

 

 

 

 

R.P. THER (1)–(3)).

 

 

 

 

 

 

 

 

 

(5) TB3 R. P. THER (1)–(3)

Short the circuit between

8.2

 

 

are opened.

terminals (1)–(3).

 

 

 

(when peripheral

 

 

 

 

equipment is not

 

 

 

 

connected between TB3

 

 

 

 

R. P. THER (1)–(3).

 

 

 

 

(6) Insufficient charging or

Check that the battery is fully

9

 

 

deterioration of the

charged. If the battery is not

10

 

 

battery.

charged, charge it. Check the

 

 

 

 

battery for deterioration and

 

 

 

 

the next replacement date.

 

 

 

 

If the battery deteriorates or

 

 

 

 

its replacement date has

 

 

 

 

passed, replace it.

 

 

 

(7) Incorrect connection of

Check that external or internal

9

 

 

the battery.

batteries are correctly

10

 

 

 

connected.

 

 

 

(8) Blown fuses F8, F9 (for

Replace the blown fuses

12.3

 

 

battery protection).

(250 V, 10 A).

 

No.5

“EMERGENCY

(1) Different model name or

Check that the model name,

Safety

 

OPERATION”

No. on the name plate.

serial number and cable length

Precautions

 

 

 

of the STP pump, the STP

4.4.2

 

 

 

control unit and the STP

 

 

 

 

connection cable are the same.

 

 

 

(2) Excessive external

After the STP pump stops

6.1.3

 

 

vibration (vibration of

thoroughly, power OFF and

 

 

 

100 ￿m p-p or more).

remove external vibration so

 

 

 

 

as not to transmit it to the STP

 

 

 

 

pump.

 

15-7