STP-H600/H1000 Series Instruction Manual

8.2DC I/O TB3 Terminal Block

This is a terminal block for a remote input/output signal. Use it in accordance with Table

8.4and Figure 8.4. Terminals excluding the HEATING terminal function in either case of the REMOTE/MANUAL operation.

The screw for the terminal is M3*1.

Four abbreviations are used in the following tables:

 

N.O

 

 

:Normal Open

N.C :Normal Close

 

IN

 

 

:Input Terminal

OUT :Output Terminal.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 8.4 DC I/O TB3 Terminal Block

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Terminal

 

 

Description

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Terminal for connecting the thermal switch to protect the

 

 

 

 

 

 

 

 

 

(1)

 

auxiliary pump.

 

R.P N.

 

 

 

 

 

 

When the normal close type thermal switch for protecting

 

 

 

 

 

 

 

the auxiliary pump is connected between this terminal,

 

CIN

THER

 

 

 

 

 

 

short the circuit between I/O TB2 R.PUMP terminal (3)-(4)

 

 

 

 

 

 

 

and stop the auxiliary pump.

 

 

 

 

 

 

 

 

 

 

 

 

(3)

 

The "FAILURE" lamp lights.

 

 

 

 

 

 

 

 

 

 

 

 

 

When this function is not used, short the circuit between (3)-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(4) (shorted at delivery).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Terminal for baking heater remote control.

(2)

 

 

 

 

 

 

 

 

 

 

 

 

When the circuit between this terminal is shorted through

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N.OIN

HEATING

 

the "REMOTE" operation, the voltage for the baking heater

 

 

 

 

 

 

 

 

 

 

 

is output between I/O TB2 HEATER terminal (1)-(2) during

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

rated operation.

(4)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Terminal for inputting the rotation INHIBIT signal. When

 

 

 

 

 

 

 

 

 

 

 

(5)

 

the terminal is set to open, the STP pump does not rotate

 

 

 

 

 

 

 

 

 

 

 

despite the STP pump operation.

 

N.CIN

INHIBIT

 

 

 

 

 

 

And only the auxiliary pump start voltage is output to I/O

 

 

 

 

 

 

 

TB2 R.PUMP terminal ((3)-(4)).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When this function is not used, short the circuit between (5)-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(7) (shorted at delivery).

 

 

 

 

 

 

 

 

 

(7)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Terminal for the emergency vent. valve operation signal.

(10)

 

 

 

 

L.VALVE

(9)

This terminal is closed while the emergency vent. valve is

 

 

 

 

opened.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N.O OUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(6)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ground

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(8)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Not-used

*1 : JIS

8-6