STP-H600/H1000 Series Instruction Manual

5.6Remote Operation

Slide the "MANUAL/REMOTE" changeover switch on the STP control unit rear panel to "REMOTE."

5.6.1Powering ON

Switch ON the breaker on the STP control unit rear panel,

(to prevent incorrect operation, a metal fitting is attached to the breaker. Loosen the screw, lift the metal fitting and secure it ).

The magnetic bearing functions and the rotor levitates (POWER ON state). The "POWER ON/OFF" switch built-in lamp lights.

5.6.2Starting/Stopping the STP Pump

There are three methods of starting/stopping the STP pump. Use one of them.

Table 5.1 Starting/Stopping the STP Pump During Remote Operation

Method

Terminal

Starting the Pump

Stopping the Pump

Block

 

 

 

1

TB5

1) Short the circuit between (3) -

Open the circuit

 

 

(2) (STOP IN).

between (3) - (2) (STOP

 

 

2) Short the circuit between (2) -

IN).

 

 

(1) (START IN) for 0.3 seconds

 

 

 

or more.

 

 

 

However, when inputting the

 

 

 

signal simultaneously with

 

 

 

switching ON the breaker on

 

 

 

the rear panel, continue to

 

 

 

short the circuit for 5 seconds

 

 

 

or more.

 

2

TB5

Short the circuit between (1) - (3)

Open the circuit

 

 

(START/STOP IN).

between (3) - (1)

 

 

 

(START/STOP IN).

3

TB6

Input 25 to 250 V AC or 24 to 48 V

Open the circuit

 

 

DC between (2) - (1) (START/STOP

between (2) - (1)

 

 

IN).

(START/STOP IN).

 

 

With DC, the (+) side can be

 

 

 

connected to either (2) or (1).

 

5.6.3Starting the STP Pump After Stopping

Perform the rotation procedures to reaccelerate the STP pump. See Section 5.6.2, "Starting/Stopping the STP Pump."

The STP pump can be reaccelerated even while the STP pump is stopping.

5-5

Page 58
Image 58
Seiko Instruments MT-17E-003-D 5.6Remote Operation, 5.6.1Powering ON, 5.6.2Starting/Stopping the STP Pump