5.6Remote Operation
Slide the "MANUAL/REMOTE" changeover switch on the STP control unit rear panel to "REMOTE."
5.6.1Powering ON
Switch ON the breaker on the STP control unit rear panel,
(to prevent incorrect operation, a metal fitting is attached to the breaker. Loosen the screw, lift the metal fitting and secure it ).
The magnetic bearing functions and the rotor levitates (POWER ON state). The "POWER ON/OFF" switch
5.6.2Starting/Stopping the STP Pump
There are three methods of starting/stopping the STP pump. Use one of them.
Table 5.1 Starting/Stopping the STP Pump During Remote Operation
Method | Terminal | Starting the Pump | Stopping the Pump | |
Block | ||||
|
|
| ||
1 | TB5 | 1) Short the circuit between (3) - | Open the circuit | |
|
| (2) (STOP IN). | between (3) - (2) (STOP | |
|
| 2) Short the circuit between (2) - | IN). | |
|
| (1) (START IN) for 0.3 seconds |
| |
|
| or more. |
| |
|
| However, when inputting the |
| |
|
| signal simultaneously with |
| |
|
| switching ON the breaker on |
| |
|
| the rear panel, continue to |
| |
|
| short the circuit for 5 seconds |
| |
|
| or more. |
| |
2 | TB5 | Short the circuit between (1) - (3) | Open the circuit | |
|
| (START/STOP IN). | between (3) - (1) | |
|
|
| (START/STOP IN). | |
3 | TB6 | Input 25 to 250 V AC or 24 to 48 V | Open the circuit | |
|
| DC between (2) - (1) (START/STOP | between (2) - (1) | |
|
| IN). | (START/STOP IN). | |
|
| With DC, the (+) side can be |
| |
|
| connected to either (2) or (1). |
|
5.6.3Starting the STP Pump After Stopping
Perform the rotation procedures to reaccelerate the STP pump. See Section 5.6.2, "Starting/Stopping the STP Pump."
The STP pump can be reaccelerated even while the STP pump is stopping.