3.3How to Install the STP Pump

STP-H600/H1000 Series Instruction Manual

3.3How to Install the STP Pump

Install the STP pump to the vacuum equipment as shown in Figure 3.2.

 

 

The STP pump can be

 

Vacuum Equipment

 

 

installed through the damper

 

 

 

 

 

Inlet Port

 

 

 

(optional accessory).

 

 

Inlet Port Flange

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Vacuum Valve

To Power Supply Power Cable Emergency

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STP Pump

 

Dry Pump,etc.

 

 

Vent.Valve

 

 

 

Emergency

 

 

 

 

 

 

 

Auxiliary

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pump

Panel

 

Vent. Valve Cable

 

 

 

 

 

 

 

STP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Front

Control Unit STP Connection Cable

 

 

 

 

 

 

Outlet Port

 

Motor Connection Cable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ground Cable

 

 

 

 

 

 

 

Leg

Remote Cable

To Vacuum Equipment Control Circuit

Figure 3.2 Installation of the STP Pump to the Vacuum Equipment

Manual backgroundManual background CAUTIONManual background

Chlorine or fluorine system gases can be used in chemical specific pumps (type C). When you use gases including alkaline metals, but excluding Li, gases including Ga, Hg, In, or Sn, or HBr, contact Seiko Instruments.

NEVER use corrosive gases (chlorine, fluorine, or other system gases) in the STP-H600/H1000 pump or other models without anti-corrosion treatment (see Section 1.1, "Usable Gases").

When you use the STP pump in a place subjected to exposure to radiation, contact Seiko Instruments.

DO NOT open the STP pump through the flange to atmospheric air while the STP pump is running.

If atmospheric air flows into the STP pump, it may not function normally.

Depending upon the type of the auxiliary pump used, atmospheric air may reverse flow into the STP pump when the auxiliary pump stops. Attach a vacuum valve to the middle of the piping between the STP pump outlet port flange and the auxiliary pump, and close the vacuum valve when the auxiliary pump stops.

3-5

Page 27
Image 27
Seiko Instruments MT-17E-003-D instruction manual 3.3How to Install the STP Pump