Seiko Instruments MT-17E-003-D Procedu, Lamp Being Lit, Probable Causes, Countermeasures, RE No

Models: MT-17E-003-D

1 124
Download 124 pages 22.02 Kb
Page 110
Image 110

STP-H600/H1000 Series Instruction Manual

Table 15.5 Troubleshooting When Any of Abnormality/Error Warning Lamp Lights

PROCEDU

Lamp Being Lit

Probable Causes

Countermeasures

Referred

RE No.

 

 

 

Section

No.1

“OVER

(1) Incorrect connection of

Connect cables correctly.

4.4.2

 

TEMPERATURE”

the STP connection

If the cable connector is not

 

 

“FAILURE”

cable.

locked, lock it securely.

 

 

“EMERGENCY

 

 

 

 

OPERATION”

 

 

 

No.2

“OVER

(1) Overheat inside the STP

 

 

 

TEMPERATURE”

pump:

 

 

 

“FAILURE”

1) Overheat during

1) Set the temperature of the

7.3

 

 

baking.

baking heater to 120 ￿C or

 

 

 

2) Insufficient cooling.

lower.

7.2

 

 

2) At the time of baking or gas

 

 

 

suction, always cool the STP

 

 

 

3) Leakage of the

pump.

3.3.4

 

 

3) Check the piping system for

 

 

vacuum piping

leakage.

 

 

 

system.

 

5.2

 

 

4) Repetitious start/stop

4) Repetitious start/stop

 

 

operations.

operations may cause the

 

 

 

 

STP pump to overheat.

 

No.3

“BATTERY

(1) Power failure.

Check a power failure.

6.1.1

 

OPERATION”

 

 

 

 

“EMERGENCY

 

 

 

 

OPERATION”

 

 

 

No.4

“FAILURE”

(1) Incorrect connection of

Connect the cable correctly. If

4.4.2

 

 

the emergency vent.

the cable connector is not

 

 

 

valve cable.

locked, lock it securely.

 

 

 

(2) When the rated operation

 

6.1.4

 

 

does not start approx. 15

 

 

 

 

minutes after start

 

 

 

 

operation is executed:

 

 

 

 

When the

 

 

 

 

ACCELERATION state

 

 

 

 

continues for approx. 15

 

 

 

 

minutes:

1) Use the maximum working

3.3.4

 

 

1) Insufficient pressure at

 

 

the inlet port.

pressure or less.

16.1

 

 

2) Insufficient pressure at

2) Use the allowable backing

3.3.4

 

 

the outlet port.

pressure or less.

16.1

 

 

3) Leakage of the piping

3) Check the piping system for

3.3.4

 

 

system.

leakage.

3.3

 

 

4) Failure of the auxiliary

4) Check the auxiliary pump

 

 

pump.

for its capacity and START

16.1

 

 

 

state.

 

 

 

 

(use a pump with the

 

 

 

 

recommended capacity or

 

 

 

 

more).

4.4.2

 

 

5) Incorrect connection of

5) Connect the cable correctly.

 

 

the motor connection

If the cable is not locked,

 

 

 

cable.

lock it securely.

 

15-6

Page 110
Image 110
Seiko Instruments MT-17E-003-D Procedu, Lamp Being Lit, Probable Causes, Countermeasures, Referred, RE No, Section