Seiko Instruments MT-17E-003-D instruction manual Abnormality

Models: MT-17E-003-D

1 124
Download 124 pages 22.02 Kb
Page 68
Image 68
Manual background

STP-H600/H1000 Series Instruction Manual

Abnormality

Operation state of safety

Lamp indication

REMOTE

State of the STP pump

Probable causes of an

/Error

functions

output signal

abnormality/error

Overheating

Temperature of the STP

"OVER

"ALARM"

The STP pump stops.

Abnormal baking temperature.

inside the

pump motor: 110 or

TEMPERATURE"

 

 

Insufficient cooling.

STP pump

more

"FAILURE"

 

 

Leakage of vacuum equipment

 

 

 

 

 

or piping.

 

 

 

 

 

Continuous repetition of

 

 

 

 

 

startup/shutdown.

 

 

 

 

 

Extraordinarily-high

 

 

 

 

 

temperature.

Overheating

90or more in the STP

"FAILURE"

"ALARM"

The STP pump stops.

Failure in the cooling fan.

inside the

control unit and heat sink.

 

 

 

Shielding of cooling air inlet

STP control

 

 

 

 

and ventilation port.

unit

 

 

 

 

Extraordinarily-high

 

 

 

 

 

temperature

Overspeed

Rotational speed of the STP

"FAILURE"

"ALARM"

The STP pump stops.

Failure in the inverter.

 

pump:37,500 rpm or more.

 

 

 

 

Abnormal

Battery terminal voltage:

"FAILURE" *2

"ALARM"

The STP pump does not start.

Incorrect connection of the

battery

42 V or lower*1

("EMERGENCY

 

 

battery.

terminal

 

OPERATION" or

 

 

Battery capacity became less

voltage

 

"BATTERY NG"

 

 

than operating limit value.

 

 

lamp may light)

 

 

 

*1 : When capacity of battery is less than operating limit value largely, there may be the thing that battery abnormal detection does not operate.

*2 : When the power is turned on again, this lamp goes out.

6-9

Page 68
Image 68
Seiko Instruments MT-17E-003-D instruction manual Abnormality