STP-H600/H1000 Series Instruction Manual

No.

Symptom

 

Probable Causes

 

Countermeasures

Referred

 

 

Section

 

 

 

 

 

 

 

2

The STP pump

(1)

Baking temperature

Set the temperature to

7.3

 

stops.

 

overheat.

120 or lower.

 

 

 

(2)

Insufficient cooling.

Always cool during baking

7.2

 

 

 

 

 

or gas suction.

 

 

 

(3)

Power failure.

Check if power has failed.

6.1.1

 

 

(4)

Over vibration

After the STP pump stops

6.1.3

 

 

(5)

(vibration 100 ￿m p-p

thoroughly, power OFF

 

 

 

 

or more).

and remove external

 

 

 

 

 

 

vibration so as not to

 

 

 

 

 

 

transmit it to the STP

 

 

 

 

 

 

pump.

 

 

 

(6)

Overheating inside the

Remove heat source or

 

 

 

 

STP control unit.

shield for air route.

 

 

 

 

1)

Ambient

1)

Keep ambient

4.2.1

 

 

 

 

temperature

 

temperature 40 ￿C or

 

 

 

 

2)

Insufficient

 

lower.

4.2.2

 

 

 

2)

Remove obstacles at

 

 

 

 

cooling of the STP

 

the ventilation port.

12.2

 

 

 

 

control unit.

 

If the air cooling fan

 

 

 

 

 

 

is broken, contact

 

 

 

 

 

 

 

Seiko Seiki.

 

 

 

(7)

Improper pressure of

Use the maximum

3.3.4

 

 

 

the inlet port.

working pressure or less.

16.1

 

 

(8)

Improper pressure of

Use the allowable backing

3.3.4

 

 

 

the outlet port.

pressure or less.

16.1

 

 

 

 

 

 

3

Rotor falls during

(1)

Failure of the STP

Inspection is needed.

4.4.2

 

operation.

 

connection cable.

Contact Seiko Seiki.

 

 

 

 

 

 

 

 

 

(2)

Failure of the battery.

Inspection is needed.

9

 

 

 

 

 

Contact Seiko Seiki.

10

 

 

 

 

 

 

4

Ultimate pressure

(1)

Failure of the auxiliary

Check the auxiliary pump

3.3

 

is insufficient.

 

pump.

for its capacity and

16.1

 

 

 

 

 

START state.

 

 

 

 

 

 

(use a pump with the

 

 

 

 

 

 

recommended capacity or

 

 

 

 

 

 

more).

 

 

 

(2)

Leakage of the piping

Check the piping system.

3.3.4

 

 

 

system.

 

 

 

 

 

(3)

Residual molecule.

If the main composition is

7.3

 

 

 

 

 

H2 or H2O, perform

 

 

 

 

 

 

baking; if it is other than

 

 

 

 

 

 

H2 or H2O, clean the

 

 

 

 

 

 

inside of the vacuum

 

 

 

 

 

 

equipment

12.4

 

 

 

 

 

(If gases remain inside

 

 

 

 

 

the STP pump, contact

 

 

 

 

 

 

Seiko Seiki when it is

 

 

 

 

 

 

needed to be cleaned).

 

15-4