3.3.1Cleaning the Seal

STP-H600/H1000 Series Instruction Manual

Manual backgroundNOTICEManual backgroundManual backgroundManual background

The STP pump cannot be used with the outlet port open to atmospheric air. Always use the auxiliary pump (dry pump or similar one).

Use an auxiliary pump with a pumping speed of 240 L/min or more.

Depending upon the type of the auxiliary pump used, oil vapor may contaminate the inside of the STP pump. Some oil viscosity could cause a malfunction when there is a strong reverse flow of oil.

Take the following measures to ensure the correct flow of oil:

￿￿Attach a vacuum valve to the middle of the piping between the STP pump outlet port flange and the auxiliary pump.

￿￿Attach an absorption trap adjacent to the vacuum valve.

3.3.1Cleaning the Seal

Inspect the seals of inlet and outlet port flanges for dirt or oil spots before installing the STP pump in the vacuum equipment.

Take the following measures for cleaning the seals:

￿￿Clean off with a pure gas.

￿￿Wipe with proper solvent (such as alcohol).

Manual backgroundManual background CAUTION Manual background

A splinter shield is attached to the inlet port flange to prevent foreign particles from falling into the STP pump.

Always leave the splinter shield attached during operation.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

The splinter shield cannot perfectly prevent foreign materials from falling into the STP pump.

DO NOT install the STP pump in such a manner that foreign materials can easily fall into it (for example, Si wafers or samples are positioned above the STP pump). If installing the STP pump in such a manner, always attach a shield plate with sufficient conductance above the STP pump to prevent foreign materials from falling into it. Foreign materials falling into the STP pump through the splinter shield may result in product damage.

Be careful not to scratch the flange of the STP pump.

Check whether or not there are scratches on the surface, before installing the STP pump.

3-6