3.3.2STP Pump Installation Positions

STP-H600/H1000 Series Instruction Manual

3.3.2STP Pump Installation Positions

The STP pump can be installed vertically, horizontally, upside-down and slanted.

Upside-down

Horizontal

STP Pump

Slanted

Vertical

Figure 3.3 STP Pump Installation Positions

When installing the STP pump in a horizontal or slanted position, it is recommended to install it so that the direction of the outlet port is on a vertical or horizontal plane in the direction of the gravity.

This makes it possible to reduce the load on the magnetic bearing and the heat generated by the STP pump.

Outlet Port

Outlet Port

Outlet Port

Outlet Port

Direction of Gravity

Figure 3.4 Positions of the Outlet Port on the Horizontally or Slanted Installed STP Pump