Seiko Instruments MT-17E-003-D instruction manual 2Disconnect battery internal connector

Models: MT-17E-003-D

1 124
Download 124 pages 22.02 Kb
Page 88
Image 88
2)Disconnect battery internal connector (6).

STP-H600/H1000 Series Instruction Manual

9.5.2How to Replace the Battery in the Battery Case

1)Remove front plate (5) from battery case (4) by unscrewing 4 M2 screws.

2)Disconnect battery internal connector (6).

3)Remove inside plate(10) from battery case by unscrewing 4 M2 screws and pull out the battery from the case (it can be pulled out toward you).

4)Replace the four worn-out batteries with new ones, the side, from which the cord projects, up.

5)Attach inside plate(10).

6)Connect battery internal connector (6) and attach front plate (5) to the battery case (Be careful not to catch the cable).

Manual backgroundManual background WARNINGManual background

When covering the battery case, be sure not to catch the battery cable in the battery case. Failure to do so may shortcircuit the battery and result in a leakage, production of fire or explosion.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

New batteries and battery cases are supplied by Seiko Seiki.

9-5