<Operation after a Power Recovery>

STP-H600/H1000 Series Instruction Manual

<Operation after a Power Recovery>

1)MANUAL Operation

The STP pump continues decelerating even after a power recovery. Press the "START" switch to reaccelerate the STP pump.

2)REMOTE Operation

￿￿When the START signal is input to the REMOTE input terminal at a power recovery to reaccelerate the STP pump.

￿￿When the START signal is not input to the REMOTE input terminal after the power recovery, the STP pump continues the BRAKE operation.

Manual backgroundNOTICEManual backgroundManual backgroundManual background

Establish a sequence so that power can be supplied to the STP control unit immediately after a power recovery.

Table 6.1 shows the states of lamps and the REMOTE output signal at a power failure.

Also, Table 6.2 shows operations of the STP pump after a power recovery.

Table 6.1 State of Lamps and REMOTE Output Signal at Power Failure

 

 

 

LED lamps

REMOTE output

 

Rotational speed

 

signal (I/O TB2)

Power failure time

 

 

“BATTERY OPERATION”

 

 

(rpm)

Power ON

ALARM

 

 

“EMERGENCY

signal

signal

 

 

OPERATION”

 

 

 

 

Approx. 2 sec. or more

Less than 2000

 

Lights

ON

ON

2000 or less

 

Goes out

OFF

OFF

 

 

Less than approx. 2 sec.

 

Does not detect power failure.

 

Table 6.2 Operation of the STP Pump after Power Recovery

 

 

 

 

 

 

 

MANUAL operation

REMOTE operation

 

Power failure time

STP pump operation

“START” REMOTE

STP pump operation

after power recovery

signal input after

after power recovery

 

 

 

 

power recovery

 

 

Approx. 2 sec. or more

Deceleration/Stop

 

Yes

Reacceleration

 

No

Deceleration/Stop

 

 

 

Less than approx. 2 sec.

 

 

Continues as before.

 

 

6-2

Page 61