Tutorials
AWG610 Arbitrary Waveform Generator User Manual 2-79
Do the following steps to load and run the sequence files:
1. Push the Waveform/Sequence bottom button.
2. Push the Load... side button.
3. Select mainseq.seq from the file list in the dialog box.
4. Push the OK side button.
If there is an error in the sequence descriptions, the instrument displays a
message and stops reading the files. Errors may occur when you use infinite
repeats in a subsequence.
. The AWG610 Arbitrary Waveform Generator reads all related sequence
files and waveform files at this time. If the instrument cannot read or find a
sequence file, it displays an error message. Make sure that you entered the
sequence and subsequence file names exactly as they appear in the file lists.
Remember that file names are case sensitive.
Do the following steps to load and run the sequence files:
1. Push the RUN button.
The RUN LED is on.
2. Push the CH 1 OUT button near the CH1 connector.
The CH1 LED is on.
When the subsequence subseq.seq is called, the AWG610 Arbitrary Waveform
Generator waits for a trigger event. The message Waiting is displayed in the
current run status area when the instrument is waiting for a trigger. The
instrument is waiting because line 1 of the main sequence is waiting for a trigger
before outputting the waveforms on that line.
. The instrument has a function that automatically provides trigger signals
at user-defined intervals. If the instrument does not wait for you to press the
Force Trigger button before executing the sequence table, you will need to
disable the automatic trigger signal. Refer to page 3–40 for information on how
to disable automatic trigger signals.
3. Push the FORCE TRIGGER button on the front panel to generate a trigger
event.
Load and Run theSequence FilesRun the Sequence Files