Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual B-47
Figure B-21: Waveformwhile all ground disclosure switches are open
b. Close the SW1 of the ground closure to generate an event signal on the
EVENT IN connector pin 0.
c. Verify that the oscilloscope displays the waveform as shown in Figure
B–22 and that the waveform is about half the amplitude as that shown in
Figure B–21.
Figure B-22: Waveformoutput when the SW1 is closed
d. Open SW1 of the ground closure to degenerate the event signal.
e. Verify that the oscilloscope displays the waveform in Figure B–21.