Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual B-49
Figure B-24: Waveformoutput when the SW3 is closed
c. Open SW3 of the ground closure to degenerate the event signal.
d. Verify that the oscilloscope displays the waveform in Figure B–21.
10. Check the EVENT IN connector pin 3 input by doing the following
substeps:
a. Close the SW4 of the ground closure to generate an event signal on the
EVENT IN connector pin 3.
b. Verify that the oscilloscope displays the waveform shown in Figure
B–25.
Figure B-25: Waveformoutput when SW4 is closed