Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual B-17
Operating Mode Tests
The following procedures verify the operation of the Cont, Triggered and Gated
modes.
. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT ( CH1 or CH1 ) LED is off.
If the other OUTPUT LED is on, push the CH1 or CH1 OUT button to turn off
the output.
Equipment
required
A50WSMAĂcoaxial cable,ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter and an
oscilloscopeĂ(TDS700).
Prerequisites The AWGArbitrary Waveform Generator must meet the prerequisites
listed onpage B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W coaxial cable to connect the AWG610 Arbitrary Waveform
Generator CH1 output connector to the oscilloscope CH1 input connector
(see Figure B–5).
OscilloscopeĂ(TDS700)
50 WSMAĂcoaxial cable
AWG610Arbitrary Waveform Generator
SMA(Fe)ĆBNC(Ma)
adapter

Figure B-5: Cont mode initial test hookup
2. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1 scale . ................. 0.2V/div
CH1 input impedance . ......... 50W
Horizontal
Sweep ..................... 200ns/div

Check Cont Mode