Appendix B: Performance Verification

AWG610 Arbitrary Waveform Generator User Manual B-21

Waveform
output CH1
Trigger
signal CH2

Figure B-7: Relationship between trigger signal andwaveform output

Equipment
required
Two50ĂWBNCĂcoaxial cables, aĂ50ĂWSMAĂcoaxial cable,
ĂaĂSMA(Fe)ĆBNC(Ma)Ăadapter,ĂaĂBNCĆTĂ(maleto 2 females) adapter, a
function generator,and an oscilloscopeĂ(TDS700).
Prerequisites The AWG610Arbitrary Waveform Generator must meet the
prerequisites listedon page B-8.

1. Set the oscilloscope controls as follows:

Vertical......................... CH1andCH2
CH1 and CH2 coupling . ........ DC
CH1scale .................. 0.5V/div
CH2scale .................. 2V/div
CH1 input impedance . ......... 50W
CH2 input impedance . ......... 1MW
Horizontal
Sweep ..................... 20ms/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto

2. Set the function generator (AFG310) controls as follows:

Check Gated Mode