Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual 
Horizontal
Sweep ..................... 2ms/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive
Level ...................... 0V
Mode . ..................... Auto
4. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls and select the waveform file:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
5. Push HORIZONTAL MENU (front-panel)Clock (side).
6. Push 1, 0, 0 and M (SHIFT+7) keys in this order or turn the general pur-
pose knob to set the clock rate to 100 MS/s.
7. Push the RUN button.
The LEDs above the RUN button and CH1 output connectors are on.
. Always perform the marker level measurements after the level has
stabilized. The marker level measurements do not include the overshoot or
undershoot.
8. Follow the substeps below to set the Marker1 controls and check the mark-
er output level :
a. Push VERTICAL MENU (front-panel)Marker... (side)Marker1
Low Level (side).
b. Push -, 1, . , 1 and ENTER keys in this order or turn the general pur-
pose knob to set the marker1 low level to -1.1 V.
c. Verify that the reading for the low level on the oscilloscope screen is
within the range between -1.255 V and -0.945 V.
d. Push Marker1 High Level (side).
e. Push 3, . , 0 and ENTER keys in this order or turn the general pur-
pose knob to set the marker1 high level to 3.0 V.