The Sequence Editor3-112 AWG610 Arbitrary Waveform Generator User ManualTable3-33: Sequence table columns
Column Description
Sequencefile name Remainsblank if you have not saved the sequence once after opening a
new file.
Line Sequence linenumber. It is assigned automatically here as a result of
the addition or deletion of a line.
CH1 Specifiesthe waveform (.WFM or .PAT) or sequence file to output on
CH 1 for that line of the sequence table.
A sequence file may be specified for an output file. Youcan only nest
sequence files one level.
The waveform file name cannot contain a drive or directory name. The
sequence file and all waveform files called must be accessible at the
same directory level. If the waveform file name fields for CH1 is blank
on a sequence line, or the instrument cannot locate a specified file, the
instrument displays an error message and aborts loading the sequence
file. Remember that file names are case sensitive.
The data length of each waveform file used for the sequence process
must be a multiple of 8 from 512 points to 8.1 M points. For sequence
output, the total of data length of the waveforms must not exceed 8.1 M
points.
Repeat Count Specifies the number of repeats. You may specify any integer from 1 to
65536, or select the keyword Infinity.The Infinity setting is neglected in
a nested sequence file (subsequence).
WaitTrigger Causes the instrument to wait for a trigger event before outputting the
waveform(s) on the specified sequence table line. Validvalues are On
and Off (blank). WaitTrigger functionality is only valid when the Run
Mode is set to Enhanced. This setting is neglected in the subsequence.
Goto One Specifieswhether control jumps to the head of the sequence table after
outputting. Validvalues are On or Off (blank). Goto One functionality is
only valid when the Run Mode is set to Enhanced. This setting is
neglected in the subsequence.
Note that, for the last line of the sequence table, this setting is always
ON independently of the setting.
Logic Jump Specifies the sequence table to jump to a specified line depending on
signal values on the EVENT IN connector.You may specify Next (go to
the next line) or Off (blank) as well as specify the sequence line number
for the destination. SelectingOff means that the line that is currently
being edited is set as a jump address. For example, when an event
occurs during theĂoutput ofĂthe waveform set in the line 5 with jump off,
the waveform in the line 5 is output again from the top. This field
remains gray if the Jump Mode is set to Tableor Software. This setting
is neglected in the subsequence or when the jump mode is set to
Software.