The Sequence Editor
AWG610 Arbitrary Waveform Generator User Manual 3-121
2
1
0
Twointernal
clock cycles
The instrument reads this state in an unstable
transitionperiod. This may cause an erroneous action.
3
0
The instrument reads this state after
the transition has completed.
3
2
1
STROBE
Strobe On
Strobe Off
Figure 3-28: Event signal timing and strobe
Software jump can be performed only with the command using a GPIB or
Ethernet interface. When you specify a line number as an argument in the
command line, the control in the currently loaded sequence file will jump to the
specified line.
To perform a software jump, the mode must be set in the loaded sequence file.
This can be set in the sequence editor by pushing:
Jump Mode (bottom)Software (side)
For more detail, refer to the description of the AWGControl:EVENt:SOFT-
ware[:IMMediate] in the AWG610 Arbitrary Waveform Generator Programmer
Manual.
The sequence is processed by the instrument hardware. The nested sequence,
however, is expanded into the sequence memory by the instrument firmware.
Software JumpLimitations on UsingSequences