Appendix B: Performance Verification
B-50 AWG610 Arbitrary Waveform Generator User Manual
c. Open SW4 of the ground closure to degenerate the event signal.
d. Verify that the oscilloscope displays the waveform in Figure B–21.
11. Retain the test hookup and control settings.
Use the test hookup and oscilloscope settings from previous check.
1. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls and select the sequence file:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Load the PT_STROB.SEQ file.
Refer to Loading Files on page B–10 for file loading procedures.
c. Push SETUP (front-panel)Run Mode (bottom)Enhanced (side) to
set the run mode to enhanced.
2. Push the RUN and CH1 OUT buttons.
The LEDs above the RUN button and CH1 output connector are on.
3. Check the EVENT IN connector strobe pin input by doing the following
substeps:
a. Verify that the waveform being displayed on the oscilloscope is shown in
Figure B–26.
Figure B-26: Initial waveformoutput
Check Strobe Input