Appendix B: Performance Verification
 AWG610 Arbitrary Waveform Generator User Manual
f. Verify that the reading for the high level on the oscilloscope screen is
within the range between 2.75 V and 3.25 V.
9. Do the following substeps to change the connection to the MARKER1
output:
a. Disconnect the 50 SMA terminator from the MARKER1 connector.
b. Disconnect the 50 SMA coaxial cable from MARKER1 connec-
tor and connect it to MARKER1.
c. Connect the 50 SMA terminator to the MARKER1 connector.
10. Repeat step 8 to check the MARKER1 output.
11. Repeat steps 1 through step 10, asreplacing the MARKER1 and MARKER1
with the MARKER2 and MARKER2, to check the MARKER2 and MARK-
ER2 outputs.
12. Disconnect the oscilloscope.