Appendix B: Performance Verification
B-58 AWG610 Arbitrary Waveform Generator User Manual
Marker Output Tests
These procedures check the accuracy of the AWG610 Arbitrary Waveform
Generator marker output level.
Equipment
required
A 50ĂWSMA coaxial cable, anĂSMA(Fe)ĆBNC(Ma) adaptĆ
er,ĂaĂ50ĂSMAĂterminator,Ăandan oscilloscope.
Prerequisites The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listedon page B-8.
. Connect a 50SMA terminator to the inverted marker output connector
during the marker output tests.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use an SMA coaxial cable and an SMA-to-BNC adapter to connect the
AWG610 Arbitrary Waveform Generator MARKER1 connector to the
oscilloscope CH1 input connector (see Figure B–31).
2. Connect a 50 SMA terminator to the MARKER1 connector.

AWG610Arbitrary Waveform Generator
50 WSMA coaxial cable
50 SMA terminator
OscilloscopeĂ(TDS700)
SMA(Fe)ĆBNC(Ma)
adapter
Figure B-31: Marker output initial test hookup
3. Set the oscilloscope controls as follows:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1 scale . ................. 1V/div
CH1 input impedance . ......... 50W
CH1offset .................. 0V